VAPO-ALPHA
Temperature compensated |
International color-coding |
Compatible with any machine |
Interlocking system with selectatec mount |
Technical specification
Dimension
Selectatec : 126mm (W) × 192mm (D) × 230mm (H)
Cage : 135mm (W) × 200mm (D) × 230mm (H)
Operation Environment
Temperature : 15 ~ 35 ℃
Humidity : 30 ~ 75 %
Agent capacity
Charged : 220mL of free volatile anesthetic agent
Charging : 300mL(nominal) to charge by vaporizers
Calibration
The dials of the Vapor-ALPHA vaporizers for enflurane, isoflurane, and halothane are calibrated to 6% v/v in steps of 0.2% up to 1%, and in steps of 0.5% from 1% to 5% at 21℃ using O₂at 5L/min as the carrier gas

Calibration of Sevoflurane
The dial of the Vapor-ALPHA vaporizer for sevoflurane is calibrated to 7% v/v
Dial is marked in steps of 0.2% up to 1% v/v, and in steps of 1% between 1% v/v and 7%
Flow rate
Vaporizer are designed to provide consistent output throghout clinical flow range from 200mL/min to 15L/min
Vaporizer type
Agent : Halothan, enflurane, Isoflurane, sevofluran
Filler type : Top filler, Quik filler(sevoflurane)
Mounting system : Cagemount, selectatec(with interlock)
Operating
Temperature : 18℃ to 35℃
Humidity : 30% to 75% relative humidity